Artificial lung ventilator for patients and hospitals for professional use. Combining the key components required for complex pulmonary ventilators together with our knowledge in Telemetry which allows us to deliver data remotely and help save lives all over the world.

The information contained in this website does not constitute a binding description of DPV 100 or its functionalities. The DPV 100 is at continuous development stage, therefore, any information provided in this website may be subject to change at Teltonika Telemedic’s sole discretion.

intuitive-control-user-interface

Intuitive control & user Interface

intergrated-cloud-connectivity

Integrated cloud connectivity

up-to-4-hours-operating-time

Up to 4 hours operating time

integrated-turbine

Integrated Turbine

incasive-and-non-incasive-modes

Invasive and Non-invasive modes

emergency-alarms

Emergency alarms

Description icon Description
Features icon Features
Downloads icon Downloads

Description

Artificial lung ventilator for patients and hospitals for professional use. Combining the key components required for complex pulmonary ventilators together with our knowledge in Telemetry which allows us to deliver data remotely and help save lives all over the world.

Features

HARDWARE
TechnologyGSM / HSPA / LTE
GNSSA-GPS, GLONASS, GALILEO, BDS
Input voltage range100 - 220 V AC
Internal back-up battery10 Ah Lithium Batteries*
Dimensions400 mm x 344 mm x 160 mm
WeightTo be defined (expected up to 10kg)
Integrated BuzzerFor alarm notifications in case of emergency
Bluetooth4.0 BLE Optional (depending on requirements)
Flash memory16 / 32 MB (To be defined)
Firmware updatesOptional cloud-based solution
Interface1 x RS232
1 x USB
1 x Emergency Call button (optional)
MicrocontrollerSTM High-performance Arm® Cortex®-M4 32-bit RISC core
DISPLAY
Resolution1280 x 800 pixels
Contrast ratio16:10 ratio (~189 ppi density)
SUPPORTED VENTILATION MODES
CPAPContinuous positive airway pressure
PC-SIMVPressure Controlled Synchronized Intermittent Mandatory Ventilation
BiPAPTwo level positive airway pressure
PC-CMVPressure Controlled Continuous Mandatory Ventilation
ALARMS
Low VTRange: 30 to 2000ml
Resolution: 10 ml
Default: 300 ml
(Lower threshold for breath cycle volume)
High VTRange: 80 to 3000ml
Resolution: 10 ml
Default: 2000 ml
(Upper threshold for breath cycle volume)
High RRRange: 10 to 70 bpm
Resolution: 1 bpm
Default: OFF
Ambient air pressure sensorIf the ambient air pressure sensor malfunctions, notification is displayed on the screen and sound alarm is being activated
Inhale/Exhale pressure sensorIn case of malfunction in inhale/exhale pressure sensors the notification is displayed on the screen and sound alarm is being activated
Oxygen flow sensorIf the ventilator detects oxygen flow sensor failure, it still can work, but FiO2 parameter is not allowed to changed. Notification is displayed on the screen and sound alarm is being activated
Inspiratory/Expiratory air flow sensorIf the sensor has failed while the ventilator was working, ventilator still works according to valves control commands which were previously set. In case of failure notification is displayed on the screen and sound alarm is being activated
Oxygen proportional valveIf an oxygen proportional valve failure is detected. Notification is displayed on the screen and sound alarm is being activated
PEEP controlIf the PEEP control solenoid failed, ventilator may still work without PEEP pressure maintenance.
Notification is displayed on the screen and sound alarm is being activated
Electricity shutdownIn case when there is unexpected electricity shut down, the ventilator can still operate up to 2-4 hours from internal batteries. Notification is displayed on the screen and sound alarm is being activated
GENERAL SETTINGS
*VALUES ARE SUBJECT TO CHANGE
PIPRange: 0 to 20 cmH2O
Resolution: 1 cmH2O
Min. VTRange: 30 to 2000ml
Resolution: 10 ml
Max. VTRange: 80 to 3000ml
Resolution: 10 ml
Min. Ins. TimeRange: 0.1 to 2.8 s
Resolution: 0.1 s
Max. Ins. TimeRange: 0.8 to 3 s
Resolution: 0.1 s
Min. FiO2Range: 18 to 90%
Resolution: 1%
Max. FiO2Range: 30 to 100%
Resolution: 1%
RtotResolution: 1 bpm
CERTIFICATION
*IN PROGRESS
RegulatoryCE, E-mark
ISO16750-2:2012
ISO 80601-2-84:2018
LST EN ISO 80601-2-12:2011
EN 794-3;2002+A2;2009
SECURITY
1Upon turning on the ventilator all system, sensors and valves functionality are being checked. If malfunction has been detected, notification is displayed on the screen and sound alarm is being activated.
2In case of electricity shutdown, internal batteries will allow to operate from 2 to 4 hours. It is designed in such way that even if a short electricity shutdown occurs, it would not affect any operations of ventilator.
3Additional air pressure, air flow and oxygen sensors double check the pressure of supplied air and oxygen mixture as well as volume and concertation. Even if one of the primary valves or sensors fails – ventilator ensures that patient would be supplied with oxygen.
4DPV100 always performs self-check procedure and checks the parameters. In case of failure of any sensor, valve or other component will result alert notification and sound alarm will be activated.